A medical-grade professional product as a body temperature guard
Model No.: BHC-601 cold and hot surgery room, intensive care unit, medical treatment for burns and scalds
● Current overload protection for machine system
● Current overload protection for control circuit system
● Independent mechanical temperature protection switch
● Two-stage power-off protection for over high and over low temperature
● Microcomputer water temperature sensor
● Surface of the outer box is coated by powder coating paint
● Trolley handle and glove box equipped
● Four-wheel design with a diagonal brake
● No electrical drainage system required
●Outward appearance
Dimensions | : | 55cm (W) X87cm (H) X 38cm (D) Weight: 43kg | |
Mattress available | : | LVC-201 54cm (W) X 70cm (D) X 1cm (Thickness) | |
LVC-202 60cm (W) X 162cm (D) X 1cm (Thickness) | |||
LVC-203 54cm (W) X 162cm (D) X 1cm (Thickness) |
●Control system
1. | Control range : Water temperature 4°C〜42°C | |
2. | Display accuracy : Water temperature ± 0.8°C | |
3. | Display range : Water temperature 0°C〜50°C | |
4. | Temperature setting : Water temperature 0.1°C | |
5. | Display mode : Three-digit display |
●Heating and refrigeration system
Compressor : 1/4 HP.R134a with eco- friendly refrigerant Heater : 750W |
●Circulation system
The total capacity of the water tank and water mattress during normal operation is about 5L | |
Flow range : 32 ~ 48GPH |
●Electrical system
Power specification : BHC-601 110V/60Hz 8Amps Max | |
BHC-601B 220V/50Hz 8Amps Max | |
Leakage current: <100uA Power cord: Medical grade |
●Safety system
1. Maximum high temperature control setting: 42°C ± 1°C | |
2. First-step high temperature protection: 45°C | |
3. Second-step high temperature protection: 46°C | |
4. Independent temperature protection switch: 50°C | |
5. Minimum low temperature control setting: 4°C ± 1°C | |
6. First-step low temperature protection: 3°C ± 1°C | |
7. Second-step low temperature protection: 2°C ± 1°C |
●Environments for use and storage
1. Ambient temperature range for use: 10°C〜40°C | |
2. Ambient humidity range for operation: 10%〜95% RH | |
3. Ambient temperature range for storage: –10°C〜+ 60°C | |
4. Ambient humidity range for storage: 10%〜95% RH |
Additional accessories:
1. |
Red and white quick connector 3-meter silicone connecting tube |
![]() |
2. | Mattresses (TPU material) available: Dimensions of LVC-201 54cm (W) X 70cm (D) X 1cm (Thickness) Dimensions of LVC-202 60cm (W) X 162cm (D) X 1cm (Thickness) Dimensions of LVC-203 54cm (W) X 162cm (D) X 1cm (Thickness) |
![]() ![]() |