4F., NO.23. LN.200, Lide st., ZHONG DIST., NEW TAIPEI CITY 235 ,TAIWAN(R.O.C)
LOGO

Hyper-Hypothermia Water System for Patient Temperature Management


Model No. : BHC-301A
分享到

A medical-grade professional product as a body temperature guard

Model No.: BHC-301A hot and cold type anesthesiology department, emergency room, surgery room

● Current overload protection for machine system

● Current overload protection for control circuit system

● Independent mechanical temperature protection switch

● Two-stage power-off protection for over high and over low temperature

● Microcomputer water temperature sensor

 

● Surface of the outer box is coated by powder coating paint

● Trolley handle and glove box equipped

● Four-wheel design with a diagonal brake

● No electrical drainage system required

Outward appearance

  Dimensions 55cm (W) X87cm (H) X 38cm (D)
  Weight 43kg
  Mattress available LVC-201 54cm (W) X   70cm (D) X 1cm (Thickness)
      LVC-202 60cm (W) X 162cm (D) X 1cm (Thickness)
      LVC-203 54cm (W) X 162cm (D) X 1cm (Thickness)

Control system

  1. Control range :Water temperature 4°C42°C    Body temperature 30°C40°C
  2. Display accuracy : Water temperature ± 0.8°C    Body temperature ± 0.6 °C
  3. Display range : Water temperature 0°C50 °C  Body temperature 10°C44°C
  4. Temperature setting : Water temperature 0.1°C   Body temperature 0.1°C
  5. Display mode: Three-digit display
  6. Patient temperature sensor : One piece
  7. Sensor model No. : YSI 400 series

Heating and refrigeration system

  Compressor : 1/4 HP.R134a with eco- friendly refrigerant   Heater : 750W

 ●Circulation system

  The total capacity of the water tank and water mattress during normal operation is about 5L
  Flow range : 32 ~ 48GPH

Electrical system power specifications

  BHC-301A 110V/60Hz 8Amps Max
  BHC-301B 220V/50Hz 8Amps Max
  Leakage current: <100uA  Power cord: Medical grade

Safety system

  1. Maximum high temperature control setting: 42°C ± 1°C
  2. First-step high temperature protection: 45°C
  3. Second-step high temperature protection: 46°C
  4. Independent temperature protection switch: 50°C
  5. Minimum low temperature control setting: 4°C ± 1°C
  6. First-step low temperature protection: 3°C ± 1°C
  7. Second-step low temperature protection: 2°C ± 1°C

Environments for use and storage

  1. Ambient temperature range for use : 10°C〜40°C
  2. Ambient humidity range for operation : 10%〜95% RH
  3. Ambient temperature range for storage : –10°C〜+ 60°C
  4. Ambient humidity range for storage : 10%〜95% RH

Additional accessories:

1.

Red and white quick connector

3-meter silicone connecting tube

 

2. Mattresses (TPU material) available:
Dimensions of LVC-201 54cm (W) X 70cm (D) X 1cm (Thickness)
Dimensions of LVC-202 60cm (W) X 162cm (D) X 1cm (Thickness)
Dimensions of LVC-203 54cm (W) X 162cm (D) X 1cm (Thickness)
3 Temperature sensor :YSI  400

YSL  400